HBPM Protocol
HBPM Protocol
HBPM Protocol
Protocol
Introduction
Home Blood Pressure Monitoring (HBPM) is being increasingly used in many health systems
worldwide and is well accepted by patients. A recent survey undertaken in the UK found that
around 30% of patients with hypertension monitor their own blood pressure at home. In other
countries, this figure stands at over 70%. To realise the full value of HBPM, however, it is important
that HBPM is carried out in a way that is consistent with the current evidence base and therefore in
line with this protocol.
All healthcare professionals (e.g. doctors, nurses, healthcare assistants, pharmacists) using HBPM
with their patients should be adequately trained and have their performance reviewed periodically.
*Please note that the guidance below frequently makes reference to patients with hypertension.
However, there will be a group of patients who do not meet the diagnostic criteria for hypertension,
but who have blood pressure too high for their clinical condition (e.g. CKD, type 2 diabetes). In
such cases, HBPM should be used as per the guidance for patients with confirmed or suspected
hypertension.
o
o
Possible white coat hypertension (blood pressure in the hypertensive range when taken
in a clinical setting, but not when taken at home)
Possible white coat effect in patients with hypertension (a difference of more than
20/10mmHg between clinic blood pressure and average home blood pressure, but where
home readings are still in the hypertensive range. Such patients are at risk of receiving
more blood pressure medication than they need)
Possible masked hypertension (blood pressure in the hypertensive range when taken at
home, but not when taken in a clinical setting)
Unusual variability of blood pressure.
o
o
o
o
o
o
o
1.1.3. Long term monitoring of patients on stable treatment for hypertension (i.e. hypertensive
patients who have well controlled blood pressure).
1.2. HBPM performed with most automated devices is contraindicated in patients with pulse
irregularities, such as atrial fibrillation (AF) [see manufacturers instructions for further
information]. Prior to introducing HBPM, therefore, in patients not known to have an irregular
pulse, it may be appropriate to check for any pulse irregularities (rate and rhythm). Some HBPM
devices will do this automatically.
1.3. HBPM should not be considered universally contraindicated in some patient groups such as
those with motor disabilities or cognitive impairments. These patients may be able to undertake
HBPM with additional support, either from a trained healthcare professional or family member.
monitor should not be used for clinical purposes and it should be recommended that the
monitor is recalibrated according to the manufacturers instructions or replaced. It is also good
practice to occasionally check patients monitors against other validated devices.
2.4. If devices are loaned to patients, from an infection control point of view, it is important to ensure
that cuffs and inflatable bladders are cleaned according to the manufacturers instructions and
in compliance with local infection control procedures.
Box 1. Cuff and Inflatable Bladder Size
Indication
Width (cm)*
Length (cm)*
10-12
18-24
BHS
Guidelines
inflatable
bladder width
and length
(cm)*
12 x 18
12-13
12-16
20
23-35
35-40
42
12 x 26
12 x 40
<53
Cuff
Sizes
Small
Adult/Child
Standard Adult
Large Adult
Adult Thigh
Cuff**
Arm
circumference
(cm)*
<33
<50
<23
* The range of columns 2 and 3 are derived from recommendations from the British Hypertension
Society (BHS), European Hypertension Society (ESH) and the American Heart Association. Columns 4
and 5 are derived only from the BHS guidelines.
** Large inflatable bladders for arm circumferences over 42cm may be required.
2.5. Manufacturers instructions should be referred to for further information e.g. setting up the
monitor, troubleshooting, battery replacement, calibration, using memory/storage functions.
artery on the arms interior surface. The forearm should be supported on a firm surface and
should be level of the lower end of the breast bone.
3.3. No tight or restrictive clothing should be worn around the arm.
3.4. Patients should be seated, remain silent and be at rest for a minimum of five minutes before
taking a measurement and should not have smoked, eaten, drunk a caffeinated drink or
undertaken physical activity within the past thirty minutes. Patients should also avoid
measuring their blood pressure with a full bladder.
3.5. Measurements should be taken in silence when the patient is relaxed, with both feet flat on the
floor and their back and arm supported. Many patients automatically cross their legs, which
raises their blood pressure, so it is particularly important to emphasise the need for the patient
to uncross their legs when taking their blood pressure.
3.6.The device should then be used as per the manufacturers instructions to produce a blood
pressure reading. The manufacturers instructions may also provide further specific guidance in
relation to the appropriate use of the device.
pressure readings
o If the monitor has been loaned, the importance of caring for and returning the equipment.
4.2. Where possible, the above education and training/teaching should consist of a practical
demonstration with the patient of how to use the blood pressure monitor.
4.3. In addition to verbal and practical education and training, written information should be provided
in relation to the above points [see Patient Information Leaflet: Home Blood Pressure
Monitoring Explained].
4.4. Patients should be provided with information about how they can seek further support if
required.
o The medication taken at the time of the recordings including name of drugs, dose and time
when taken should be recorded and be available for review.
5.1.3. Long term monitoring of hypertensive patients on stable treatment for hypertension:
o For each blood pressure recording, at least two consecutive measurements should be
taken, at least one minute apart. Additional measurements should be taken where the first
two measurements taken are quite different
o Blood pressure can be taken less frequently than above, for example, a maximum of once
or twice per week, although this could be inappropriately intensive for some. Overuse of
HBPM should be avoided
o Where blood pressure has been stable for a long time, intermittent checks, 2 readings
twice daily for 4-7 consecutive days (with average recorded), every 4 to 6 months are
advised
o For patients on stable treatment, immediately prior to the patients annual hypertension
review, consider asking the patient to record their blood pressure for a course of four,
ideally seven, days in accordance with the schedule outlined in 5.1.2.
o It is important that the records of the blood pressure and all medication are made available
for any clinical review
o It is recommended that the patient brings their blood pressure machine with them to
reviews so that an appropriately trained healthcare professional can check that a validated
device and the right sized cuff are being used, and also that the device is being used in the
correct way. Equivalence with clinic blood pressure readings can also be checked.
use the same monitor and it is vital not to confuse which readings belong to each patient. It is
also good practice to clear the memory of any monitor between patients, particularly where
devices are loaned to patients.
7.
7.1. Where HBPM is used short-term for diagnostic purposes or to monitor the treatment of
hypertension, readings taken on the first day should be discarded and an average of the
remaining values- systolic and diastolic individually- taken.
7.2. Where patients have been monitoring their blood pressure at home previously (e.g. previously
for diagnostic purposes and now to monitor treatment response), it may not be appropriate to
discard the readings taken on the first day, although this issue is still a matter of debate.
7.3. If patients have taken more than two measurements at each recording, due to their first two
readings being vastly different, consider including only the last two measurements when
calculating the average.
7.4. To achieve equivalence with clinic-measured blood pressure, home blood pressure readings
should be adjusted by -5/5mmHg. For example, a blood pressure reading of 135/85 taken at
home is equivalent to a reading of 140/90 in a clinic setting.
IT IS VITALLY IMPORTANT THAT HOME BLOOD PRESSURE READINGS ARE ADJUSTED
IN THIS WAY AS FAILURE TO DO SO CAN RESULT IN AN UNDERTREATMENT AND
UNDERDIAGNOSIS OF HYPERTENSION.
7.5. Healthcare professionals should explain the interpretation of the home blood pressure
measurements provided, and their implications for management, with the patient and, where
appropriate, his or her carer.
This resource is a joint production of the NIHR Collaboration for Leadership in Applied Health
Research and Care (CLAHRC) Greater Manchester and the British Hypertension Society
Recording and coding HBPM information in the patients record (e.g. using recognised Read
codes in GP practices)
Processes for administering HBPM devices to patients
Related guidance/resources:
National Institute for Health and Clinical Excellence (NICE), (2011). The clinical management of
primary hypertension in adults: Clinical Guideline 127. NICE: London.
Parati et al., (2010). European Society for Hypertension Practice Guidelines for home blood
pressure monitoring. Journal of Human Hypertension, 24(12), pp. 779-785.
www.bloodpressureuk.org
This resource is a joint production of the NIHR Collaboration for Leadership in Applied Health
Research and Care (CLAHRC) Greater Manchester and the British Hypertension Society