Describe the bug
During Custom Preset configuration, for both Saved and Custom (single use) Presets, the Target Range uses manual input for setting the low and high value. Low value can be entered as low as 0 mg/dL and high value can be raised as high as 400mg (at least).
Given the low value is unconstrained, it is unclear if the dose will still be constrained by Glucose Safety Limit. Note: this should be investigated.
Additionally the UI should introduce picker interactions rather than manual inputs consistent to Correction Range in Therapy Settings to mitigate against use error. See below.
To Reproduce
Steps to reproduce the behavior:
Go to 'Custom Presets'
Click on '+' to create a new saved preset or 'Custom' to create a temporary preset
Scroll down to 'Target Range'
Tap into input fields, enter 0 or 400 into lower and upper bound.
Expected behavior
The UI should introduce picker interactions rather than manual inputs consistent to Correction Range in Therapy Settings to mitigate against use error.
Additionally, two recommendations to increase consistency:
Change Target Range to Correction Range across the Preset interface
Display warning message as user’s selection approaches high and low limits
Trigger a low range value warning and verify the following copy: “A value you have entered is lower than what is typically recommended for most people.”
Trigger a high range value warning and verify the following copy: “A value you have entered is higher than what is typically recommended for most people.”
Trigger a high and low range value warning and verify the following copy “Some of the values you have entered are outside of what is typically recommended for most people.”
Screenshots
The text was updated successfully, but these errors were encountered:
Describe the bug
During Custom Preset configuration, for both Saved and Custom (single use) Presets, the
Target Range
uses manual input for setting the low and high value. Low value can be entered as low as 0 mg/dL and high value can be raised as high as 400mg (at least).Given the low value is unconstrained, it is unclear if the dose will still be constrained by
Glucose Safety Limit
. Note: this should be investigated.Additionally the UI should introduce picker interactions rather than manual inputs consistent to
Correction Range
inTherapy Settings
to mitigate against use error. See below.To Reproduce
Steps to reproduce the behavior:
Expected behavior
<title></title> <style type="text/css"> p.p1 {margin: 0.0px 0.0px 0.0px 0.0px; font: 13.0px 'Helvetica Neue'; min-height: 15.0px} p.p2 {margin: 0.0px 0.0px 0.0px 0.0px; font: 13.0px 'Helvetica Neue'} p.p3 {margin: 0.0px 0.0px 0.0px 0.0px; font: 12.0px Helvetica; min-height: 14.0px} table.t1 {border-collapse: collapse} td.td1 {border-style: solid; border-width: 1.0px 1.0px 1.0px 1.0px; border-color: #9a9a9a #9a9a9a #9a9a9a #9a9a9a; padding: 1.0px 5.0px 1.0px 5.0px} </style>The UI should introduce picker interactions rather than manual inputs consistent to
Correction Range
inTherapy Settings
to mitigate against use error.Additionally, two recommendations to increase consistency:
Change
Target Range
toCorrection Range
across the Preset interfaceDisplay warning message as user’s selection approaches high and low limits